SpinCon® 45010A Specifications
|
|
Power Requirements:
Internal battery or external 120VAC |
Sampling Time:
Selectable between 5 min and 6 hours |
Power Consumption:
450-500W |
Sampling Size:
10mL |
Optional Batteries:
12V/33-amp hour sealed rechargeable battery with protective case |
Sampling Extraction:
Automatic into sample bottle - features bottle presence detection |
Flow Rate:
450 LPM |
Sampling Solutions:
Water or phosphate buffered saline (PBS) - consult factory if your application requires a custom solution |
Size:
18"x15"x8" |
User Interface:
Integral control with LCD status display; sample, rinse, sample time, and manual extract and inject controls |
Weight:
46 lbs |
Decontamination:
External surfaces can be wiped with dilute bleach solution; fluid lines and contactor can be flushed with bleach solution followed by rinses with sterile water |
Operating Temperature Range
2°-50° C (36°-120° F) |
|
OMNI™ 3000 Specifications
|
|
Power Requirements:
Internal battery or external 120VAC |
Sampling Time:
User selectable up to 6 hours |
Power Consumption:
70W max while sampling (< 2.5W Idle) |
Sampling Size:
< 9mL |
Optional Batteries:
24VDC hot-swappable nickel metal hydride |
Sampling Extraction:
Automatic into sealed disposable sample cartridge |
Flow Rate:
300 LPM with integral flow regulation |
Sampling Solutions:
Water, phosphate buffered saline (PBS), or PBS with Triton - uses water for evaporative makeup - consult factory if your application requires a custom solution |
Size:
7.5”x8.5”x16.5” |
User Interface:
Sealed touch pad controls with LCD display easily usable and viewable in personal
protective equipment |
Weight:
13 lbs (with optional battery < 21 lbs) |
Decontamination:
Alcohol or bleach - OMNI has smooth surfaces and minimal hard-to-clean areas - external surfaces can be sprayed or rinsed, fluid lines can be easily flushed, and contactor can be removed for submersion cleaning or replacement |
Operating Temperature Range
2°-50° C (36°-120° F) |
Concentration Rate:
At 1µm 120,000; at 3µm 150,000 (assuming an ambient concentration of 1,000 particles per liter of air) |